SPECIFICATIONS
- Automatic hydrogen maintenance
- (Designed to control at precisely 3% for safe operating conditions.)
- Touch Screen Controls
- Alarms for excess gas consumption
- Trouble shooting guide
- Data logging of both oxygen and hydrogen levels
- Maximize cost efficiency of the Gas Mix supply
- Includes communication cable for connection to any Coy Anaerobic Monitor
- Gas Lines and Chamber Fittings Included
When the chamber hydrogen level drops to 2.8% or lower. Valve on the Infuser opens to allow the anaerobic mix gas to enter into the chamber automatically. At the same time a one way valve opens allowing gas to slowly escape through positive pressure created by the injection of fresh gas mix. If the hydrogen levels do not reach 3.0% within 2.5 hours an alarm will trigger and the system will stop. The alarm has to be cleared prior to operating again and an on‐screen guide is provided to diagnose the reason for the alarm.
| Hydrogen Control Set Point | 3% H2 |
| Control Hysteresis | 0.1% H2 |
| Timer Alarm | 2.5 hrs after purge starts |
Instruction Manual
To avoid pressurizing the system a small valve opens allowing excess pressure to vent from the system through tubing connections.
The Anaerobic Gas Infuser also includes data logging software allowing you to chronicle your oxygen and hydrogen levels. Tracking this data is ideal for trouble shooting, minimizing waste and improving lab efficiency.